Features-
- Compact and lightweight
- One touch operation for all 12 leads
- Automatic and Manual modes
- Last patient’s ECG storage
- Paper-out indication
- Indication for Battery Status, Charging and AC/DC Operation
- Auto Base line control to optimize ECG lead printing
Measurements:
Technical Specifications- -HR(bpm) : -RR,QRS,QT,QTC,PR(intervals)
- -QT/QTC, QT/RR(%): -P/QRS/T axis
ECG
- A/D Resolution: 10 Bit
- Lead Selection: Standard 12 Leads
- Filters: Low pass filter: 35, 100Hz , High pass filter: 0.05Hz, Notch filter 50 Hz (on/off)
- Acquisition : Auto/Manual
- Memory : Last patient
- Input impedance : >100 M ohm
- Frequency Response 0.05-100Hz
- CMRR : 100dB
- Time Constant : 3.2 sec
- Sensitivity : 5mm/mV, 10mm/mV,20mm/mV
- Leakage Current : <10 micro Ampere
- Recorder : Thermal Printer
- Paper Speed : 12.5mm/sec., 25mm/sec., 50mm/sec.
- Recording Paper : Thermo Sensitive 50mm (Rolled)
- Printing Labels : Patient Information Templates, Leads, Calibration, Print Speed, Filter, Sensitivity
- Print –out Format : Single Channel
- Line Power supply : ~220-240v, 50Hz ± 10%
- Power Supply Operation : AC/DC with on/off line Battery Charging
- Battery : ___12V/1.2AH
- Charging Time : 2Hours
- Dimensions : 227mm X168mm X55mm
- Weight : 1.45 Kg approx.
- Type of Protection : Class II A
- Degree of Protection : Type BF
- Temperature : 0-40˚C
- Non-condensing humidity : 0- 80%
- Temperature : 0-35°C
- Component : Quantity
- ECG Patient Cable : 01
- Adaptor 19V, 3.4A, AC-DC : 01
- Power Cord : 01
- Fuse T 1.5 A : 01
- Earthing Wire : 01(Optional)
- Bulb Electrode : 06
- Limb Electrode : 04
- Conducting Tube Jelly : 01
- Thermal Printer Paper Roll : 01
- User Manual : 01